Z-0154-06
Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers...
Medrad, Inc. / initiated 2005-10-10 / Terminated
Medrad Inc began a recall of Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX on . The reason FDA records is "the heat maintainer, an accessory, may overheat and deform the disposable syringe". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0154-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX
- Product code
- DXT Injector And Syringe, Angiographic
- Reason for recall
- the heat maintainer, an accessory, may overheat and deform the disposable syringe
- FDA root cause
- Other
- Action
- The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/12/05 via certified mail. The notice provided instructions to apply warning labels and disconnect the heat maintainer until a field upgrade kit is available. In case where the heat maintainer is a necessity for specific procedures, the customer is being instructed to observe certain precautions.
- Quantity in commerce
- 4362 units
- Distribution
- The products were shipped to medical facilities nationwide. The products were also shipped to Canada, Isreal, Italy, Denmark, Netherlands, Norway, Great Britain, France, Poland, Germany, Spain, Finland, Belgium, Sweden, Russia, Austraila, Cuba, Mexico, Peru, Hong Kong, Korea, Mylasia, Thialand, Taiwan, Singapore, India, China, Turkey, Saudia Arabia, and Japan
- Product codes and lots
- all serial numbers