Z-0154-06

Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers...

Medrad, Inc. / initiated 2005-10-10 / Terminated

Medrad Inc began a recall of Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX on . The reason FDA records is "the heat maintainer, an accessory, may overheat and deform the disposable syringe". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0154-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medrad, Inc.
Product
Stellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX
Product code
DXT Injector And Syringe, Angiographic
Reason for recall
the heat maintainer, an accessory, may overheat and deform the disposable syringe
FDA root cause
Other
Action
The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/12/05 via certified mail. The notice provided instructions to apply warning labels and disconnect the heat maintainer until a field upgrade kit is available. In case where the heat maintainer is a necessity for specific procedures, the customer is being instructed to observe certain precautions.
Quantity in commerce
4362 units
Distribution
The products were shipped to medical facilities nationwide. The products were also shipped to Canada, Isreal, Italy, Denmark, Netherlands, Norway, Great Britain, France, Poland, Germany, Spain, Finland, Belgium, Sweden, Russia, Austraila, Cuba, Mexico, Peru, Hong Kong, Korea, Mylasia, Thialand, Taiwan, Singapore, India, China, Turkey, Saudia Arabia, and Japan
Product codes and lots
all serial numbers