Z-0137-06

Vitros ECi Immunodiagnostic System, sold as Cat. #863 3893; Cat. #192 2814 (Vitros ECiQ), and Cat. #680 1059...

Ortho-Clinical Diagnostics, Inc. / initiated 2005-09-29 / Terminated

Ortho-Clinical Diagnostics began a recall of Vitros ECi Immunodiagnostic System, sold as Cat. #863 3893; Cat. #192 2814 (Vitros ECiQ), and Cat. #680 1059 (recertified Vitros ECi) when using Vitros... on . The reason FDA records is "When using Vitro ECi Immunodiagnostic System with Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs, results of the 1st of two required re-test replicates may be negtively biased". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0137-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Vitros ECi Immunodiagnostic System, sold as Cat. #863 3893; Cat. #192 2814 (Vitros ECiQ), and Cat. #680 1059 (recertified Vitros ECi) when using Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs. NOTE: These 3 catalog numbers of Vitros ECi represent the original unit, an updated unit with cosmetic changes only, and refurbished units.
Product code
KHO Fluorometer, For Clinical Use
Reason for recall
When using Vitro ECi Immunodiagnostic System with Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs, results of the 1st of two required re-test replicates may be negtively biased.
FDA root cause
Other
Action
Letters dated 9/29/2005 issued to all domestic customers of the VITROS ECi/ECiQ systems and to Ortho/J&J foreign affiliates. Customers instructed to immediately discontinue using reflex processing to the VITROS Anti-HBc IgM and VITROS Anti-HAV reagent packs.
Quantity in commerce
3,507 Vitros ECi units distributed
Distribution
Nationwide and worldwide distribution of the VITROS ECi Immunodiagnostic systems.
Product codes and lots
All Vitros ECi units currently on the market (software version 3.1 or lower).