Z-0097-06

CeeOn Heparin SurfaceModifie (HSM) PMMA One-Piece Anterior Chamber IOLs (Model: HSM65)

Advanced Medical Optics, Inc. / initiated 2005-10-05 / Terminated / root cause: packaging

Advanced Medical Optics, Inc. began a recall of CeeOn Heparin SurfaceModifie (HSM) PMMA One-Piece Anterior Chamber IOLs (Model: HSM65) on . The reason FDA records is "A small opening (channel or tunnel) in the seal of the outer Tyvek pouch which contains the lens case, in which the sealed intraocular lens is packaged may not be adequately sealed, therefore the potentially compromised...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0097-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
CeeOn Heparin SurfaceModifie (HSM) PMMA One-Piece Anterior Chamber IOLs (Model: HSM65)
Product code
HQL Intraocular Lens
Reason for recall
A small opening (channel or tunnel) in the seal of the outer Tyvek pouch which contains the lens case, in which the sealed intraocular lens is packaged may not be adequately sealed, therefore the potentially compromised outer Tyvek pouch may not be sterile and may potentially contaminate an otherwise sterile field during surgery.
Root cause tag
packaging
FDA root cause
Other
Action
A product notification letter was mailed to all US domestic customers via overnight carrier (Federal Express) beginning on October 5, 2005. A global notification letter will be translated into the local languages by the international AMO affiliates and mailed to international customers as soon as a listing of all customers is tabulated.
Quantity in commerce
45 (4 in US)
Distribution
Nationwide, Austria, Belgium, Switzerland, Germany, Denmark, France, UK, Guadelope, Hong Kong, Croatia, India, Italy, Netherlands, and Vietnam
Product codes and lots
CeeOn¿ One-Piece PMMA HSM65, 2 lenses  PMMA HSM HSM65+22,0  8054720508 8054730508