Z-0173-06

VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester...

Ortho-Clinical Diagnostics, Inc. / initiated 2005-10-12 / Terminated

Ortho-Clinical Diagnostics began a recall of VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626 on . The reason FDA records is "Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0173-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626
Product code
LEH Radioimmunoassay, Vancomycin
Reason for recall
Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days.
FDA root cause
Other
Action
Letters dated 10/12/2005 issued to end users in the U.S. Customers advised they may continue to use the VANC Reagent if their laboratory processes QC fluids according to protocol. A return reply form is included. Foreign affiliates were notified via e-mail on 10/12/2005.
Quantity in commerce
784 boxes x 6 packs x 50 tests
Distribution
Product shipped to medical facilities (end-users) in the U.S. and to 9 International J&J affiliates in Australia, Canada, Caribbean, Chile, France, Philippines, Portugal, Taiwan, UK/Northern Europe, Israel, Saudi Arabia and Venezuela.
Product codes and lots
01-5919, Exp. 5/31/2005 (EXPIRED):  01-6032, Exp. 9/30/2005 (EXPIRED); and 02-6284, Exp. 1/28/2006