Z-0173-06
VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester...
Ortho-Clinical Diagnostics, Inc. / initiated 2005-10-12 / Terminated
Ortho-Clinical Diagnostics began a recall of VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626 on . The reason FDA records is "Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0173-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626
- Product code
- LEH Radioimmunoassay, Vancomycin
- Reason for recall
- Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days.
- FDA root cause
- Other
- Action
- Letters dated 10/12/2005 issued to end users in the U.S. Customers advised they may continue to use the VANC Reagent if their laboratory processes QC fluids according to protocol. A return reply form is included. Foreign affiliates were notified via e-mail on 10/12/2005.
- Quantity in commerce
- 784 boxes x 6 packs x 50 tests
- Distribution
- Product shipped to medical facilities (end-users) in the U.S. and to 9 International J&J affiliates in Australia, Canada, Caribbean, Chile, France, Philippines, Portugal, Taiwan, UK/Northern Europe, Israel, Saudi Arabia and Venezuela.
- Product codes and lots
- 01-5919, Exp. 5/31/2005 (EXPIRED): 01-6032, Exp. 9/30/2005 (EXPIRED); and 02-6284, Exp. 1/28/2006