Z-0138-06

Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735

Siemens Medical Solutions USA, Inc. / initiated 2005-10-05 / Terminated

Siemens Medical Solutions USA, Inc began a recall of Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735 on . The reason FDA records is "If the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0138-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
If the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an incorrect location and presents the possibility of an inconclusive diagnosis. calculation of offset values for the breast biopsy aiming device may be incorrect
FDA root cause
Other
Action
The recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions MR034/05/S. The letter re-emphasizes the importance of printing the offset values on the worksheet prior to moving the patient table and in accordance with the User's Manual.
Quantity in commerce
16 units
Distribution
The product was shipped to medical facilities in CA, FL, IL, MD, MN, NM, OH, OR, RI, and WI.
Product codes and lots
Model number 5728735