Z-0893-06
Life Pulse High Frequency Ventilator (HFV)
Cutler Campus / initiated 2005-04-04 / Terminated
Bunnell, Inc. began a recall of Life Pulse High Frequency Ventilator (HFV) on . The reason FDA records is "Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0893-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cutler Campus
- Product
- Life Pulse High Frequency Ventilator (HFV)
- Product code
- LSZ Ventilator, High Frequency
- Reason for recall
- Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing the solenoid in the valve to function sporadically.
- FDA root cause
- Other
- Action
- Notified consignees by letter on 04/04/2005.
- Quantity in commerce
- 297 units
- Distribution
- Nationwide, Military medical center, Canada, and Czech Republic
- Product codes and lots
- Serial numbers 2585 and below.