Z-0201-06
PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565
Ferno-Washington Inc / initiated 2003-03-28 / Terminated
Ferno-Washington Inc began a recall of PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565 on . The reason FDA records is "The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0201-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ferno-Washington Inc
- Product
- PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565.
- Product code
- FPO Stretcher, Wheeled
- Reason for recall
- The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.
- FDA root cause
- Other
- Action
- The firm went out to customers who had complained of a 'fold' and installed the kit on the stretcher that had experienced the fold and on all other PROflexx stretchers owned by the customer. These customers had received a recall letter, dated March 28, 2003. For customers that had not complained of the stretchers 'folding', the firm sent a letter dated June 24, 2003, to all distributors requesting the distributor provide the firm with a list of customers that had received the PROflex stretchers. Once the firm had a list of all customers, they sent them a letter offering them the kit as an enhancement that they could purchase to help reduce the possibility of false locking.
- Quantity in commerce
- 2188 total stretchers
- Distribution
- The product was distributed nationwide to 27 states: TN, MA, VT, IL, KY, NY, SC, WI, TX, MO, NH, CO, GA, NC, CA, VA, IA, PA, CT, OH, NJ, ID, WV, AL, AR, ME and IN.
- Product codes and lots
- All stretchers containing a serial number beginning with the letter ''L'' and followed by a sequence of six numbers.