Z-0857-06
Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU)
GE Healthcare / initiated 2005-11-16 / Terminated
GE Healthcare began a recall of Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU) on . The reason FDA records is "Alarm sound-It is possible that the audible alarms from the speaker from the Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU) machine may dampen or disappear as a result of normal handling of the anesthesia machine...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0857-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU)
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Alarm sound-It is possible that the audible alarms from the speaker from the Datex-Ohmeda S/5 Anesthesia Delivery Unit (ADU) machine may dampen or disappear as a result of normal handling of the anesthesia machine. Visual alarms on the ADU are unaffected by this condition. The dampening or absence of audible alarms on the ADU may result in a delay in treatment.
- FDA root cause
- Other
- Action
- Customers were sent a letter dated November 16, 2005, informing them of the possible alarm failure and giving them short term and long term solutions. Short term, clinicians are to: 1) check the audio alarm prior to use; and 2) use the ADU in conjunction with functioning anesthetic agent monitor. In the long term a GE Healthcare representative will contact the facility and schedule a visit to implement the corrective action at no charge.
- Quantity in commerce
- 113---(USA-61--International-52)
- Distribution
- Worldwide---USA-States of AZ, CA, IA, MI, TX, and WI ----International- Australia, Austria, Belgium, Finland, Germany, Israel, Italy, Jordan, Malaysia, Netherlands, Norway, Portugal, Slovakia, South Africa, and Turkey.
- Product codes and lots
- 6025108 thru 6025111, 6025114 thru 6025117, 6025119, 6025120, 6025122 thru 6025128, 6025130 thru 6025132, 6025134, 6025136, 6025137, 6035737, 6035738, 6035740 thru 6035745, 6035747, 6035751, 6035757, 6035758, 6035760, 6035762, 6035765, 6035772 thru 6035774, 6035784, 6048756 thru 6048761, 6048764, 6048847, 6048849, 6048853 thru 6048856 and 40300736