Z-0534-06
Sigma implantable pulse generators (IPGs) dual Chamber Pacemaker, models SD203, SD303
Medtronic, Inc. / initiated 2005-11-29 / Terminated
Medtronic Inc. Cardiac Rhythm Managment began a recall of Sigma implantable pulse generators (IPGs) dual Chamber Pacemaker, models SD203, SD303 on . The reason FDA records is "An issue exists with a specific subset of Sigma series pacemakers that may fail due to separation of interconnect wires from the hybrid circuit. This failure mechanism may present clinically as loss of rate response...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0534-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Sigma implantable pulse generators (IPGs) dual Chamber Pacemaker, models SD203, SD303
- Product code
- DXY Implantable Pacemaker Pulse-Generator
- Reason for recall
- An issue exists with a specific subset of Sigma series pacemakers that may fail due to separation of interconnect wires from the hybrid circuit. This failure mechanism may present clinically as loss of rate response, premature battery depletion, intermittent or total loss of telemetry, or not output. There have been no reported patient injuries or deaths due to this issue.
- FDA root cause
- Other
- Action
- Press Release was issued 11/29/05. Hospital/physician letters sent via registered mail on 11/29/05 for delivery on 12/01/05. Unimplanted devices are being retrieved. The Physician letter describes the issue, root cause, the probability of reoccurrence and provides recommendations. A list of affected serial numbers that the physician is following or has implanted is attached to the letter. The letter also provides medtronic's website in which regular updates on the ongoing actual performance are listed in their Product Performance Report. Medtronic is providing a replacement device should the Physician and patient elect to replace the affected implanted device.
- Quantity in commerce
- 3,588
- Distribution
- Nationwide. and to OUS including: Europe, Australia, Canada, Hong Kong and Greater China, Japan, Latin America
- Product codes and lots
- Not all serial numbers within these ranges are affected. The specific serial numbers of affected devices may be found online at http://SigmaSNList.medtronic.com. Model SD203: serial no. PJD182918H - PJD191025H, PJD432660H - PJD449026H, PJD674196S - PJD690881S, PJD731443S - PJD743447S. Model SD303: serial no. PJD183969H - PJD186285H, PJD187584H - PJD187735H, PJD190274H, PJD190305H, PJD190308H, PJD195395H, PJD195444H, PJD441326 - PJD448026H, PJD674200S - PJD681640S, PJD691135S, PJD691231S, PJD729895S - PJD742209S.