AquaPak Sterile Water for Inhalation, USP, 760 mL blow-molded plastic bottle for use with a respiratory gas...

Teleflex Medical / initiated 2005-12-01 / Terminated / root cause: sterility

Teleflex Medical began a recall of AquaPak Sterile Water for Inhalation, USP, 760 mL blow-molded plastic bottle for use with a respiratory gas humidifier; 10 bottles per case; manufactured by... on . The reason FDA records is "Teleflex Medical determined that the distribution center shipped unreleased product which did not pass validated sterility test requirements". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0503-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
AquaPak Sterile Water for Inhalation, USP, 760 mL blow-molded plastic bottle for use with a respiratory gas humidifier; 10 bottles per case; manufactured by Hudson Respiratory Care, Inc., Temecula, CA 92589-9020 USA; catalog #037-00
Product code
BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
Reason for recall
Teleflex Medical determined that the distribution center shipped unreleased product which did not pass validated sterility test requirements.
Root cause tag
sterility
FDA root cause
Other
Action
Teleflex Medical sent recall letters dated 12/1/05 to their direct accounts on the same date via UPS overnight. The accounts were informed that the lot did not pass validated sterility test requirements, and were instructed to cease distribution and use of the lot and quarantine it for return to Teleflex Medical, Durham, NC. The wholesalers were requested to notify their customers of the recall and have them return their inventory of the lot to the direct account for consolidation and return to Teleflex. Questions were directed to Dan Daily at 1-800-334-9751, ext. 4948.
Quantity in commerce
1050 units
Distribution
Colorado, Nebraska, Oregon, Minnesota, California, Pennsylvania, Michigan, Louisiana, Oklahoma and Texas.
Product codes and lots
catalog number 037-00, lot 150150, exp. 2009-10