Z-0345-06

Introducer Set. Labeled as ''Luer-Lock Peel-Away Introducer Set *** Model: LLP-7 *** Size: 7F (2.33mm)''...

Oscor Inc. / initiated 2005-09-13 / Terminated

Oscor Inc. began a recall of Introducer Set. Labeled as ''Luer-Lock Peel-Away Introducer Set *** Model: LLP-7 *** Size: 7F (2.33mm)'' Catalog # 06607 (Oscor) and 666089-201 (Guidant) on . The reason FDA records is "Sheath size is larger than dilator size causing insertion difficulties". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0345-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Oscor Inc.
Product
Introducer Set. Labeled as ''Luer-Lock Peel-Away Introducer Set *** Model: LLP-7 *** Size: 7F (2.33mm)'' Catalog # 06607 (Oscor) and 666089-201 (Guidant)
Product code
DYB Introducer, Catheter
Reason for recall
Sheath size is larger than dilator size causing insertion difficulties.
FDA root cause
Other
Action
The firm will issue a recall letter to its consignees informing them of the recall. The letter will ask the consignees to return the devices from affected lot numbers for replacement. The returned products will be individually inspected.
Quantity in commerce
6985
Distribution
Distributed throughout the United States and to several foreign countries including China, Japan, Germany and the Phillippines.
Product codes and lots
BBraun Lot # 60710907 and 60710908 Oscor Lot # OL09342-1 and OL9645-01