Z-0280-06

Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein...

Boston Scientific Corporation / initiated 2005-12-02 / Terminated / root cause: manufacturing process

Boston Scientific began a recall of Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein Introduction Only. Order No. 50-501. Sterile EO. Single... on . The reason FDA records is "There have been reports of detachment at the bond between the carrier capsule and the outer sheath of the Greenfield Vena Cava Filters with 12 Fr Femoral introducer Systems manufactured before Marcyh 10, 2004. If the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0280-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein Introduction Only. Order No. 50-501. Sterile EO. Single Use Only. Read Instructions for Use before using this device. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. Manufacturer: Boston Scientific Cork Ltd., Business & Technology Park, Model Farm Road, Cork, Ireland. The Stainless Steel Greenfield Vena Cava Filter with 12 Fr. / 4,0 mm introducer System is a permanently implanted stainless steel device designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. The Stainless Steel Greenfield Vena Cava Filter comes preloaded in a 12 Fr. / 4,0 mm jugular or femoral introducer catheter.
Product code
DTK Filter, Intravascular, Cardiovascular
Reason for recall
There have been reports of detachment at the bond between the carrier capsule and the outer sheath of the Greenfield Vena Cava Filters with 12 Fr Femoral introducer Systems manufactured before Marcyh 10, 2004. If the capsule should detach during an implantation procedure, there is a risk of cardiac and pulmonary embolization.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Recall packages started going out 12/02/05 via Fed Ex and was completed 12/03/05. A letter was directed to several persons within each hospital . The letter describes the issues, the affected lot numbers and the complaints that Boston Scientific has received. Affected devices are to be segregated for return back to Boston Scientific. A 'Reply Verification Tracking Form' is asked to be returned back to Boston Scientific.
Quantity in commerce
18,000
Distribution
nationwide (except ND); Virgin Islands, OUS to include: united Arab Emirates, Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Chile, China, Colombia, Germany, Spain, France, Ireland, Great Britain, Israel, India, Italy, Jamaica, Jordan, Japan, Korea (South Republic), Kuwait, Lebanon, Mexico, Malaysia, Netherlands, New Zealand, Panama, Philippines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Taiwan, Venezuela
Product codes and lots
all codes of product manufactured before March 10, 2004. Catalog no. 50-501 with lot/batch # between 5145758 and 6387904