Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424...
Medtronic Inc, Neurological & Spinal Division began a recall of Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only)... on . The reason FDA records is "A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0370-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model 7428 Kinetra Neurostimulator; Models 3387, 3389 DBS Leads; and Models 7482,7495 DBS Extensions. Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas of the brain. The Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via Extension and Lead. These components comprise the implantable portion of the Activa System.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate (SAR) of not more than 0.4 W/kg. The recommended maximum displayed head SAR is now 0.1 W/kg.
- FDA root cause
- Other
- Action
- The firm''s recall consists of providing revised instructions for the Activa DBS. The revision modifies the recommended SAR to use during MRI of patients who have an Activa DBS. Use of the Activa DBS is otherwise unaffected by this recall. A Dear Healthcare Provider letter and MRI Radiologist letter were sent on 11/30/05 and 12/05/05 to provide this information.
- Distribution
- Nationwide and OUS to includeCanada, Europe, Middle East, Africa, Japan, Asia-Pacific, Australia - New Zealand, Latan America
- Product codes and lots
- All serial numbers and Lot codes are affected.