Z-0361-06
Roche brand Omni S2 Blood Gas Analyzer; Catalog # 03337111001
Roche Diagnostics GmbH / initiated 2005-12-09 / Terminated
Roche Diagnostics Corp. began a recall of Roche brand Omni S2 Blood Gas Analyzer; Catalog # 03337111001 on . The reason FDA records is "Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0361-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche brand Omni S2 Blood Gas Analyzer; Catalog # 03337111001.
- Product code
- KHG Whole Blood Hemoglobin Determination
- Reason for recall
- Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.
- FDA root cause
- Other
- Action
- Consignees were notified via letter dated 12/9/05.
- Distribution
- California, Florida, Georgia, Indiana, Iowa, New Hampshire, Michigan, New York, North Carolina, North Dakota, Pennsylvania, Texas and Washington.
- Product codes and lots
- All units with Co-Oximetry (COOX) modules.