Z-0430-06

BD PhoenixSpec Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit...

Becton Dickinson & Co / initiated 2005-12-01 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of BD PhoenixSpec Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit calibrators, and labeled in part ***Becton, Dickinson and... on . The reason FDA records is "A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0430-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD PhoenixSpec Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit calibrators, and labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA***
Product code
JQX Nephelometer, For Clinical Use
Reason for recall
A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The recalling firm notified consignees by phone, fax and e-mail on 12/01/05. The notification advised that nephelometers displaying 'CAL' error messages on the LED should not be used until calibrated with a different kit lot. Consignees were requested to discard referenced lots for replacement. A response form was included in the notification.
Quantity in commerce
145 kits
Distribution
Product was sold to direct accounts nationwide and international affiliated distributors in Japan, Belize, Taiwan, Canada, Korea, Australia, and the Philippines.
Product codes and lots
Lot 5075281 Exp 2/14/06*** Lot 5115146 Exp 2/14/06