Z-0397-06

VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable...

Plus Orthopedics / initiated 2005-12-02 / Terminated / root cause: manufacturing process

Plus Orthopedics USA began a recall of VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument) on . The reason FDA records is "The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0397-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Plus Orthopedics
Product
VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument)
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The adjustment know tower on one device was found t have broken off from the body of the instrument. This did not happend during a surgery, and is belived tohave been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
Quantity in commerce
14
Distribution
Arizona, California, Colorado, Idaho, New York, Oklahoma, Texas, Utah, Wisconsin
Product codes and lots
PI.79;  PI.79/I;  PI.79/IIII;  PI.80;  PI.80/III;  PI.80/IIII;  PI.80/IIIII