VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable...
Plus Orthopedics USA began a recall of VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument) on . The reason FDA records is "The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0397-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Plus Orthopedics
- Product
- VKS Knee System; VKS Distal Femoral Resection Guide, Model Number 1214122, Revisions A and B. (Re-usable surgical instrument)
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- The adjustment know tower on one device was found t have broken off from the body of the instrument. This did not happend during a surgery, and is belived tohave been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
- Quantity in commerce
- 14
- Distribution
- Arizona, California, Colorado, Idaho, New York, Oklahoma, Texas, Utah, Wisconsin
- Product codes and lots
- PI.79; PI.79/I; PI.79/IIII; PI.80; PI.80/III; PI.80/IIII; PI.80/IIIII