Sickle-STAT Test Kit, Catalog # SC901, a qualitative test kit for the determination of the presence of...
Chembio Diagnostic System Inc began a recall of Sickle-STAT Test Kit, Catalog # SC901, a qualitative test kit for the determination of the presence of hemoglobin S in human blood --- FOR IN VITRO DIAGNOSTIC... on . The reason FDA records is "Firm received one report of a false negative with the positive control. Internal testing by the firm has shown that the kit lot may result in invalid or false negative interpretations". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0470-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Chembio Diagnostic Systems, Inc.
- Product
- Sickle-STAT Test Kit, Catalog # SC901, a qualitative test kit for the determination of the presence of hemoglobin S in human blood --- FOR IN VITRO DIAGNOSTIC USE --- FOR PROFESSIONAL USE ONLY --- Contents: 48 pre-filled Sickle-STAT reaction tubes containing sodium hydrosulfite. 1 bottle phosphate buffer containing surfactant and stabilizer, and 1 Instruction leaflet. --- manufactured by: CHEMBIO DIAGNOSTIC SYSTEMS, INC., MEDFORD, NY 11763 --- Accessories available: SC902: Sickle-STAT 5-place Observation rack, SC903: Sickle-STAT Hemoglobin S Control Set; 4 x 1 vials: 2 A/A negative and 2 A/S positive, SC904: Sickle-STAT Phosphate Buffer, 200 ml.
- Product code
- GHM Test, Sickle Cell
- Reason for recall
- Firm received one report of a false negative with the positive control. Internal testing by the firm has shown that the kit lot may result in invalid or false negative interpretations.
- FDA root cause
- Other
- Action
- Chembio phoned the consignees and followed up by faxing the recall letters (dated 12/15/05) & response forms on 12/22/05.
- Quantity in commerce
- Lot SC050305: 347 test kits, Lot SC062805: 410 test kits
- Distribution
- Medical facilities and a few distributors located nationwide, the Dominican Republic, and West Indies (St. Vincent & Grenadine)
- Product codes and lots
- Lot SC050305, exp. 31OCT06; Lot SC062805, exp. 30NOV06