Z-0539-06

Sigma implantable pulse generators (IPGs) Single Chamber Pacemaker model SVVI103, for ventricular use only

Medtronic, Inc. / initiated 2005-11-29 / Terminated

Medtronic Inc. Cardiac Rhythm Managment began a recall of Sigma implantable pulse generators (IPGs) Single Chamber Pacemaker model SVVI103, for ventricular use only on . The reason FDA records is "An issue exists with a specific subset of Sigma series pacemakers that may fail due to separation of interconnect wires from the hybrid circuit. This failure mechanism may present clinically as loss of rate response...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0539-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Sigma implantable pulse generators (IPGs) Single Chamber Pacemaker model SVVI103, for ventricular use only.
Product code
DXY Implantable Pacemaker Pulse-Generator
Reason for recall
An issue exists with a specific subset of Sigma series pacemakers that may fail due to separation of interconnect wires from the hybrid circuit. This failure mechanism may present clinically as loss of rate response, premature battery depletion, intermittent or total loss of telemetry, or not output. There have been no reported patient injuries or deaths due to this issue.
FDA root cause
Other
Action
Press Release was issued 11/29/05. Hospital/physician letters sent via registered mail on 11/29/05 for delivery on 12/01/05. Unimplanted devices are being retrieved. The Physician letter describes the issue, root cause, the probability of reoccurrence and provides recommendations. A list of affected serial numbers that the physician is following or has implanted is attached to the letter. The letter also provides medtronic's website in which regular updates on the ongoing actual performance are listed in their Product Performance Report. Medtronic is providing a replacement device should the Physician and patient elect to replace the affected implanted device.
Quantity in commerce
2,989
Distribution
Nationwide. and to OUS including: Europe, Australia, Canada, Hong Kong and Greater China, Japan, Latin America
Product codes and lots
Not all serial numbers within these ranges are affected. The specific serial numbers of affected devices may be found online at http://SigmaSNList.medtronic.com.  Model SVVI103: serial no.PJG159902H - PJG176701H, PJG432779H - PJG432895H, PJG230582M - PJG236398M, PJG677442S - PJG696014S, PJG748310S - PJG755468S, PJG747654C - PJG755171C.