Z-0398-06

ARCHITECT c8000 System Software, Version 2.11, Product List/Model #5F48-12, distributed by Abbot...

Abbott Medical / initiated 2005-12-08 / Terminated / root cause: software

Abbott Laboratories, Inc began a recall of ARCHITECT c8000 System Software, Version 2.11, Product List/Model #5F48-12, distributed by Abbot Laboratories, Irving, TX on . The reason FDA records is "System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0398-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT c8000 System Software, Version 2.11, Product List/Model #5F48-12, distributed by Abbot Laboratories, Irving, TX.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL.
Root cause tag
software
FDA root cause
Other
Action
Firm initiated recall to all consignees via letter dated 12/08/05.
Quantity in commerce
1206 units.
Distribution
Worldwide. Product distributed nationwide to direct accounts (including gov''t accounts) in: AL, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MN, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WY. Foreign distribution to the following countries: Argentina, Australia, Bermuda, Brazil, Canada, Chile, Columbia, Ecuador, El Salvador, Germany, Grand Cayman, Guatemala, Hong Kong, Jamaica, Mexico, New Zealand, Panama, Puerto Rico, Singapore, Thailand, Trinidad, Uruquay and Venezuela.
Product codes and lots
All lots/serial numbers.