Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0382-06

Roche Elecsys total PSA CalSet, catalog number 11731696322

Roche Diagnostics GmbH / initiated 2005-12-07 / Terminated

Roche Diagnostics Corp. began a recall of Roche Elecsys total PSA CalSet, catalog number 11731696322 on . The reason FDA records is "There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0382-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Elecsys total PSA CalSet, catalog number 11731696322.
Product code
LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
Reason for recall
There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
FDA root cause
Other
Action
Consignees were notified of the incompatibility via reagent bulletin dated 12/7/05.
Quantity in commerce
25,000 kits
Distribution
Nationwide.
Product codes and lots
Lot 171735.