Z-0382-06
Roche Elecsys total PSA CalSet, catalog number 11731696322
Roche Diagnostics GmbH / initiated 2005-12-07 / Terminated
Roche Diagnostics Corp. began a recall of Roche Elecsys total PSA CalSet, catalog number 11731696322 on . The reason FDA records is "There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0382-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche Elecsys total PSA CalSet, catalog number 11731696322.
- Product code
- LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
- Reason for recall
- There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
- FDA root cause
- Other
- Action
- Consignees were notified of the incompatibility via reagent bulletin dated 12/7/05.
- Quantity in commerce
- 25,000 kits
- Distribution
- Nationwide.
- Product codes and lots
- Lot 171735.