Z-0384-06
Clinical Chemistry Immunoglobulin M, List Number 1E01-20
Abbott Medical / initiated 2005-11-10 / Terminated
Abbott Laboratories Diagnostic Div began a recall of Clinical Chemistry Immunoglobulin M, List Number 1E01-20 on . The reason FDA records is "These specific protein assays are not meeting the 57-day on board stability claim". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0384-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Clinical Chemistry Immunoglobulin M, List Number 1E01-20
- Product code
- DFT Igm, Antigen, Antiserum, Control
- Reason for recall
- These specific protein assays are not meeting the 57-day on board stability claim
- FDA root cause
- Other
- Action
- A Device Correction letter identifying the changes being made to the frequency of QC monitoring for these assays, will be distributed to all AEROSET and ARCHITECT c8000 customers which utilize the IgM, Prealbumin and Apolipoprotein Al Specifc Protein assays on November 10,2005 via direct mailing. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check..
- Quantity in commerce
- 1275 (212 in US)
- Distribution
- Nationwide, plus Australia, Canada, Columbia, Germany, Guatemala, Hong Kong, mexico, & New Zealand
- Product codes and lots
- Lot Number(s): 30067HW00, 27057HW00, & 23026HW00