Z-0513-06

ARCHITECT STAT Troponin-I Reagent Kit, list 2K41-20; each 400 test kit contains 4 bottles each of...

Abbott Medical / initiated 2005-12-21 / Terminated / root cause: manufacturing process

Abbott Laboratories MPG began a recall of ARCHITECT STAT Troponin-I Reagent Kit, list 2K41-20; each 400 test kit contains 4 bottles each of microparticles, conjugate and assay diluent; Abbott... on . The reason FDA records is "There has been an increase in customer complaints observed for failure to calibrate or inability to generate patient result errors". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0513-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT STAT Troponin-I Reagent Kit, list 2K41-20; each 400 test kit contains 4 bottles each of microparticles, conjugate and assay diluent; Abbott Laboratories, Abbott Park, IL 60064
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
There has been an increase in customer complaints observed for failure to calibrate or inability to generate patient result errors.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Abbott sent recall letters dated 12/20/05 to the laboratory accounts via Federal Express Priority on 12/21/05. The accounts were informed of the increase in customer complaints for failure to calibrate or inability to generate patient result errors for the Troponin-I Reagent lot, and were requested to discontinue use and destroy any of the affected lot found in inventory for replacement.
Quantity in commerce
181 kits
Distribution
Nationwide and internationally to Canada, Australia, Germany and Taiwan
Product codes and lots
list 2K41-20, lot 95567G005, exp 6-30-2006