Z-0472-06
Uritek TC-101Urine Analyzer, catalog number TC-101
Teco Diagnostics / initiated 2005-12-22 / Terminated
Teco Diagnostics began a recall of Uritek TC-101Urine Analyzer, catalog number TC-101 on . The reason FDA records is "When reading the Nitrite parameter on the Teco Diagnostics URS-10 strips, some positive nitrite samples may be read by the machine as negative". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0472-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teco Diagnostics
- Product
- Uritek TC-101Urine Analyzer, catalog number TC-101
- Product code
- KQO Automated Urinalysis System
- Reason for recall
- When reading the Nitrite parameter on the Teco Diagnostics URS-10 strips, some positive nitrite samples may be read by the machine as negative.
- FDA root cause
- Other
- Action
- On December 22nd,2005 the firm sent the notification letter along with a replacement unit of TC-101. These letters and replacement units were send via Fed ex ground and the tracking numbers and serial numbers of the sent units are also noted. Also noted are the return authorization numbers assigned to each respective customer.
- Quantity in commerce
- 50
- Distribution
- Nationwide
- Product codes and lots
- Serial# 200509001 through 200509050