Z-0829-06
IntraStent DoubleStrut LD - Balloon Expandable Biliary Stent, Sterilized EO, ev3 4600 Nathan Lane North...
Medtronic, Inc. / initiated 2005-12-29 / Terminated / root cause: labeling and UDI
Ev3, Inc began a recall of IntraStent DoubleStrut LD - Balloon Expandable Biliary Stent, Sterilized EO, ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA. The IntraStent... on . The reason FDA records is "IntraStent product S15-36 (P/N 90-1341-02) lot 494428 was labeled as IntraStent product S15-26 (P/N 90-1341-001)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0829-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- IntraStent DoubleStrut LD - Balloon Expandable Biliary Stent, Sterilized EO, ev3 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA. The IntraStent DoubleStrut LD Biliary Stent is a balloon expandable stent. It is intended as a permanently implanted device. The stent is made from a 316L stainless steel tube cut into an open lattice design. It is designed to be crimped onto a non-compliant PTA balloon catheter that is indicated for biliary stent expansion. After mounting onto a balloon catheter, it is deployed and expanded by inflating the balloon.
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- IntraStent product S15-36 (P/N 90-1341-02) lot 494428 was labeled as IntraStent product S15-26 (P/N 90-1341-001)
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- On March 3, 2004, shipping personnel noted that IntraStent product S15-36, lot 494428, was labeled as IntraStent product S15-26. Six units from the lot were in distribution. On December 29, 2005, customers were sent a Field Action Form to record the number of used or available for return stents at their facility. The form instructs the customer to return the completed original form and any unused stents.
- Quantity in commerce
- 6 devices
- Distribution
- AR, GA, IN, MA, TN, and UT
- Product codes and lots
- Model S15-36, lot no. 494428