Z-0822-06
Haemonetics Cell Saver 5 Autogolous Blood Recovery System Model Numbr: LN02005-110-E
Haemonetics Corporation / initiated 2005-12-21 / Terminated
Haemonetics Corporation began a recall of Haemonetics Cell Saver 5 Autogolous Blood Recovery System Model Numbr: LN02005-110-E on . The reason FDA records is "Device emits radiofrequency which exceed the IEC 60601-1-2 International Standard". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0822-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Haemonetics Corporation
- Product
- Haemonetics Cell Saver 5 Autogolous Blood Recovery System Model Numbr: LN02005-110-E
- Product code
- CAC Apparatus, Autotransfusion
- Reason for recall
- Device emits radiofrequency which exceed the IEC 60601-1-2 International Standard
- FDA root cause
- Other
- Action
- Haemonetics notified accounts by Certified Letter on 12/21/05. A field correction will be schedule to replace the LCD .
- Quantity in commerce
- 3 units
- Distribution
- Nationwide
- Product codes and lots
- Serial Numbers: 94H180CD-1; 95A048CS; 95J177CS