Z-0391-06

Cook brand Zilver 635 Biliary Stent - Expanding Stent, delivery system length 80 cm, stent diameter 6.0 mm...

Cook Incorporated / initiated 2005-10-24 / Terminated / root cause: labeling and UDI

Cook, Inc. began a recall of Cook brand Zilver 635 Biliary Stent - Expanding Stent, delivery system length 80 cm, stent diameter 6.0 mm, stent length 60 mm, minimum guiding catheter 8.0... on . The reason FDA records is "The side of the box gives the size of the stent as 8 x 60 instead of the actual size of 6 x 60. The label front is correct". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0391-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook brand Zilver 635 Biliary Stent - Expanding Stent, delivery system length 80 cm, stent diameter 6.0 mm, stent length 60 mm, minimum guiding catheter 8.0 French, minimum sheath 6.0 French, recommended wire guide size .035 inch dia., sterile; Catalog # ZIB6-80-6.0-60.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
The side of the box gives the size of the stent as 8 x 60 instead of the actual size of 6 x 60. The label front is correct.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were notified of the initial recall by phone beginning on 10/24/05 and of the extension to the second product on 01/06 /06.
Quantity in commerce
9
Distribution
Connecticut, Michigan, Minnesota, Montana, Nevada, Pennsylvania, Utah, Wisconsin and Canada.
Product codes and lots
Lot 1544171.