Z-0563-06
Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific...
Cygnus, LLC / initiated 2005-02-04 / Terminated
Cygnus LLC began a recall of Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade on . The reason FDA records is "The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0563-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cygnus, LLC
- Product
- Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
- Product code
- JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
- Reason for recall
- The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
- FDA root cause
- Other
- Action
- A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
- Quantity in commerce
- 20 units
- Distribution
- All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741
- Product codes and lots
- Serial Numbers: For MTF: -- 034135 through 034144; For Cascade -- 034119 through 034128.