Centrifuge Model CF-6 with lid lock. (Block Model Number OCT-VI). Units are distributed to Bloch Scientific...
Cygnus LLC began a recall of Centrifuge Model CF-6 with lid lock. (Block Model Number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade on . The reason FDA records is "When device was packaged, too much expanding foam was added causing the lid to deform and causing the lid locking mechanism to be out of alignment". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0564-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cygnus, LLC
- Product
- Centrifuge Model CF-6 with lid lock. (Block Model Number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
- Product code
- JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
- Reason for recall
- When device was packaged, too much expanding foam was added causing the lid to deform and causing the lid locking mechanism to be out of alignment.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- A recall notice was sent to Bloch Scientific on September 8, 2005. Bloch further distributed the product to MTF, Cascade and will conduct a sub-recall
- Quantity in commerce
- 110 units
- Distribution
- All product was shipped domestically to the firm''s distributor: Block Scientific, 1170 Lincoln Ave., Unit 11, Holbrook, NY 11741
- Product codes and lots
- Serial Numbers: Labeled for MFT -- 033824; 033825; 033827 through 033840; 033842 through 033844; 033852 through 033855; 033857 through 033863; 034411 through 034415; 034417 through 034445; 08296594 through 08296608. Labeled for Cascade -- 033690; 033989; 034109; 033942 through 033956; 033958 through 033960; 034119 through 034128.