Z-0527-06
CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer
Abbott Medical / initiated 2005-12-16 / Terminated
Abbott Laboratories began a recall of CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer on . The reason FDA records is "T-Cell test results and the respective absolute counts for the CELL-DYN Sapphire instrument using CD3/4/8 reagents can be overestimated by as much as 13%". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0527-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- T-Cell test results and the respective absolute counts for the CELL-DYN Sapphire instrument using CD3/4/8 reagents can be overestimated by as much as 13%.
- FDA root cause
- Other
- Action
- The firm has sent notification letters dated Dec. 16, 2005 to all consignees requesting discontinuing of use, destruction of CD3/4/8 reagent if the lab does not plan to run on a CELL-DYN 4000 system.
- Quantity in commerce
- 871 units
- Distribution
- Nationwide and Canada, Venezuela, Germany, Hong Kong, Japan, Mexico and Singapore.
- Product codes and lots
- List Number: 07H22-01, lot numbers: 16328 and 29401