CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer

Abbott Medical / initiated 2005-12-16 / Terminated

Abbott Laboratories began a recall of CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer on . The reason FDA records is "T-Cell test results and the respective absolute counts for the CELL-DYN Sapphire instrument using CD3/4/8 reagents can be overestimated by as much as 13%". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0527-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
CELL-DYN CD 3/4/8 Assay Reagents for Sapphire Hematology Analyzer
Product code
GKZ Counter, Differential Cell
Reason for recall
T-Cell test results and the respective absolute counts for the CELL-DYN Sapphire instrument using CD3/4/8 reagents can be overestimated by as much as 13%.
FDA root cause
Other
Action
The firm has sent notification letters dated Dec. 16, 2005 to all consignees requesting discontinuing of use, destruction of CD3/4/8 reagent if the lab does not plan to run on a CELL-DYN 4000 system.
Quantity in commerce
871 units
Distribution
Nationwide and Canada, Venezuela, Germany, Hong Kong, Japan, Mexico and Singapore.
Product codes and lots
List Number: 07H22-01, lot numbers: 16328 and 29401