Z-0522-06
The product is a mobile 3-wheeled electric scooter, model HMV 250. The product contains either a red or blue...
Tung Keng Enterprise Co., Ltd. / initiated 2005-11-01 / Terminated
Tung Keng Enterprise Co., Ltd. began a recall of The product is a mobile 3-wheeled electric scooter, model HMV 250. The product contains either a red or blue base. The red base is identified with SKU... on . The reason FDA records is "The throttle spring, located in the hand throttle assembly, can fracture, resulting in the scooter not stopping when the user releases the throttle". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0522-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tung Keng Enterprise Co., Ltd.
- Product
- The product is a mobile 3-wheeled electric scooter, model HMV 250. The product contains either a red or blue base. The red base is identified with SKU #915363224. The blue base is identified with SKU #915363225. The product is distributed by the Invacare Corporation.
- Product code
- INI Vehicle, Motorized 3-Wheeled
- Reason for recall
- The throttle spring, located in the hand throttle assembly, can fracture, resulting in the scooter not stopping when the user releases the throttle.
- FDA root cause
- Other
- Action
- The recalling firm sent field correction instructions to their customers on or about 11/1/05.
- Quantity in commerce
- 8,213 total devices
- Distribution
- The product was distributed to OHL/Invacare, Rancho Technology Park Bldg. B 8950, Rancho Cucamonga, CA 91730; OHL/Invacare, 1100 Whitaker Road, Plainfield, IN 46168; Invacare Corporation, 39400 Taylor Parkway North Ridgeville, OH 44039; and Invacare, Dallas Texas Distribution Center, 3460 South Watson Rd., Arlington, TX 76014.
- Product codes and lots
- All serial numbers 04H86001 to 05A468246.