Z-0675-06
her option Cryoablation Therapy Disposable Probe, Model CU-1. Sterile. Manufactured at: American Medical...
American Medical Systems, Inc. / initiated 2006-01-13 / Terminated
American Medical Systems began a recall of her option Cryoablation Therapy Disposable Probe, Model CU-1. Sterile. Manufactured at: American Medical Systems, Inc., 10700 Bren Road West, Minnetonka, MN... on . The reason FDA records is "Overheating of the Cryoablation Therapy Disposable Probe (CU-1), could cause a burn injury to a patient or user in certain circumstances". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0675-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- American Medical Systems, Inc.
- Product
- her option Cryoablation Therapy Disposable Probe, Model CU-1. Sterile. Manufactured at: American Medical Systems, Inc., 10700 Bren Road West, Minnetonka, MN 55343 USA. The Her Option Cryoablation Therapy System consists of a Console, a Cryoprobe and a Disposable Probe.
- Product code
- MNB Device, Thermal Ablation, Endometrial
- Reason for recall
- Overheating of the Cryoablation Therapy Disposable Probe (CU-1), could cause a burn injury to a patient or user in certain circumstances.
- FDA root cause
- Other
- Action
- An Urgent Device Recall / Product Safety Advisory letter, dated January 25, 2006, was sent to all customers who have purchased Disposable Probes since 01/01/2005 or who have in their possession a Her Option console (whether customer owned or AMS owned). The letter informs customers of the issue and of the 2 lots being recalled. The letter also includes Safety Advisory Instruction which are to be posted on or near the Her Option Console. A Recall Acknowledgement Form is asked to be returned to American Medical Systems which serves as an indication that the customer read the notice, is returning affected product and have understood and posted the Safety Advisory instructions.
- Quantity in commerce
- 152
- Distribution
- Nationwide. AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NJ, NM, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI
- Product codes and lots
- Lots 431803 & 436666