Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02
Abbott Laboratories Diagnostic Div began a recall of Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02 on . The reason FDA records is "The two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0516-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- The two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two lots are now expired. The expiration date for Lot number 26674M100 will be revised from 2006-07-31 to 2006-01-31.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- A Device Correction letter identifying the changes being made to the expiration dating for lot numbers 19906M200, 20879MI00 and 26674M1O0 along with a Customer Reply form will be distributed via FedEx to all AEROSET and ARCHITECT c8000 customers which utilize the Clincial Chemistry Multiconstituent Calibrator on December 27,2005. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check.
- Quantity in commerce
- 13961
- Distribution
- Nationwide plus worldwide in Mexico,Venezuela, Chile, Brazil, San Salvador, Singapore, Korea, Trinidad, Grand Cayman, Guatemala, Columbia, Uruguay, Puerto Rico, Canada, Hong Kong, Australia, Bermuda, Jamaica, Costa Rica, Ecuador, Argentina, Dominican ReD., Germanv, Thailand, New Zealand, Bahamas, Panama, & Paraguay.
- Product codes and lots
- Lot number(s): 19906M200 (expiration date 02/28/2006), 20879MIOO (expiration date 02/28/2006), & 26674MIOO (expiration date 07/31/2006)