Baxter Epidural Anesthesia Trays that contain BD Catheter Connectors; Sterile, single use only kits...
Baxter Healthcare Corp. began a recall of Baxter Epidural Anesthesia Trays that contain BD Catheter Connectors; Sterile, single use only kits containing an epidural needle, epidural catheter, glass LOR... on . The reason FDA records is "The BD catheter connector contained in the Epidural Anesthesia Trays may not fully secure the catheter and allows the catheter to either become loose or disconnected from the connector". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0509-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Epidural Anesthesia Trays that contain BD Catheter Connectors; Sterile, single use only kits containing an epidural needle, epidural catheter, glass LOR syringe and non-latex components for the preparation and insertion of the epidural needle and catheter; Assembled by Baxter Healthcare Corporation, Deerfield, IL U.S.A.
- Product code
- CAZ Anesthesia Conduction Kit
- Reason for recall
- The BD catheter connector contained in the Epidural Anesthesia Trays may not fully secure the catheter and allows the catheter to either become loose or disconnected from the connector.
- FDA root cause
- Other
- Action
- Baxter sent Urgent Product Recall letters dated 1/10/06 to their customers, to the attention of the Director of Anesthesia, via first class mail on the same date. The accounts were informed that the BD catheter connector in the tray may not fully secure the catheter, resulting in the possible leakage or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and destroy it for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
- Quantity in commerce
- 2,200 trays
- Distribution
- Nationwide
- Product codes and lots
- product code 2T0609, lot GD822585; product code 2T2263, lot GD822874; product code 2T2367, lots GD822551 and GD822981; product code 2T2386, lot GD822452