Z-0565-06
AUDICOR 200 D Inovise Part Number 30047 (console part number 20158)
Inovise Medical, Inc. / initiated 2003-01-12 / Terminated
Inovise Medical, Inc. began a recall of AUDICOR 200 D Inovise Part Number 30047 (console part number 20158) on . The reason FDA records is "Use of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0565-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Inovise Medical, Inc.
- Product
- AUDICOR 200 D Inovise Part Number 30047 (console part number 20158).
- Product code
- DPS Electrocardiograph
- Reason for recall
- Use of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power.
- FDA root cause
- Other
- Action
- On January 12, 2006, 2 of 3 customers were notified by telephone, with follow-up email. All three customers were visited by the firm on January 16-17, 2006. Customers were instucted to not use the device in-line with a defibrillator/monitor.
- Quantity in commerce
- 28 units
- Distribution
- Product distributed to three EMS/Fire Departments located in Florida and Washington
- Product codes and lots
- Serial Numbers: 411005, 411007, 411008, 411010, 411011, 411013, 411014, 411015, 411016, 411017, 411018, 411019, 411020, 411023, 411024, 411025, 411026, 411030, 411031, 411032, 411033, 411047, 411048, 411049, 411051, 411052, 411053, 411054.