Z-0528-06
IMMULITE 2000 Total Testosterone Kit
Siemens Healthcare Diagnostics Inc. / initiated 2006-01-03 / Terminated / root cause: manufacturing process
Diagnostic Products Corp. began a recall of IMMULITE 2000 Total Testosterone Kit on . The reason FDA records is "The reagent for this kit lot (L2KTW 2,6 126) was manufactured with a raw material that can carryover and interfere with DPC Kit IMMULITE 2000 Estradiol (L2KE2) causing a false elevation in the observed estradiol...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0528-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE 2000 Total Testosterone Kit
- Product code
- CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone
- Reason for recall
- The reagent for this kit lot (L2KTW 2,6 126) was manufactured with a raw material that can carryover and interfere with DPC Kit IMMULITE 2000 Estradiol (L2KE2) causing a false elevation in the observed estradiol concentration.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- For domestic customers - the recall is extended to the domestic end user-level. Domestic customers were notified via telephone using a Phone Script (attached in Attachment 5), starting on January 5,2005. Each customer receives a fax-back form and a Technical Bulletin directly after this phone contact. For foreign consignees - the recall is extended to foreign distributors. It is the distributor''s responsibility to extend the recall action to their direct customers. Foreign consignees were notified via e-mail and/or FAX,starting on January 3, 2005.
- Quantity in commerce
- 710
- Distribution
- Nationwide and Worldwide in Pakistan, Italy, Cyprus, Turkey, Switzerland, Chile, South Africa, Greece, Mexica, Canada, Thialand, Germany, India, United Arab Emirates, Hungary, Israel, Portugal, Taiwan, Austria, Saudi Arabia, Malaysia, Hong Kong, India, Colombia, Argentina, Australia, Costa Rica, Czech Republic, Spain, France, Panama, Brazil, Netherlands, Poland, Sweden, United Kingdom, & El Salvadon.
- Product codes and lots
- CATALOG & LOT NUMBERS: L2KTW 2 Lot 126 (200-test kit) L2KTW 6 Lot 126 (600-test kit)