Accu-Chek Advantage blood glucose test strips; Reference numbers 03144917001, 03144704001, 03145263001, (sold...
Roche Diagnostics Corp. began a recall of Accu-Chek Advantage blood glucose test strips; Reference numbers 03144917001, 03144704001, 03145263001, (sold within the United States) and 12030551001... on . The reason FDA records is "There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0567-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Accu-Chek Advantage blood glucose test strips; Reference numbers 03144917001, 03144704001, 03145263001, (sold within the United States) and 12030551001, 12030578001 and 12030586001 (sold outside of the United States).
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
- FDA root cause
- Other
- Action
- Notifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.
- Distribution
- Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan and Thailand.
- Product codes and lots
- All lots.