Z-1209-06
Midland Deluxe Mat Platform with Adjustable Backrest Model C5522-30A, B, and C
Performance Health Supply / initiated 2006-01-25 / Terminated
SAMMONS PRESTON ROLYAN began a recall of Midland Deluxe Mat Platform with Adjustable Backrest Model C5522-30A, B, and C on . The reason FDA records is "There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1209-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Performance Health Supply
- Product
- Midland Deluxe Mat Platform with Adjustable Backrest Model C5522-30A, B, and C,
- Product code
- INT Plinth
- Reason for recall
- There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- FDA root cause
- Other
- Action
- Letters were sent to customers certified mail on 1/25-26/2006 instructing them to take the tables out of service until it could be inspected by a sales representative. Sales represenatives are following up with customers in person to inspect tables.
- Quantity in commerce
- 831 devices total
- Distribution
- Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV.
- Product codes and lots
- Serial number prefixess 2509 throuth 2511