Z-0577-06

Stat Profile Critical Care PLUS (CCX) Analyzer Catalog Number: 37413

Nova Biomedical Corporation / initiated 2006-01-27 / Terminated / root cause: manufacturing process

Nova Biomedical Corporation began a recall of Stat Profile Critical Care PLUS (CCX) Analyzer Catalog Number: 37413 on . The reason FDA records is "Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0577-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care PLUS (CCX) Analyzer Catalog Number: 37413
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs
Root cause tag
manufacturing process
FDA root cause
Other
Action
Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.
Quantity in commerce
408 units
Distribution
Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI
Product codes and lots
All S/N''s