Z-0599-06
Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) and...
Baxter Healthcare Corporation / initiated 2006-02-08 / Terminated
Baxter Healthcare Corp. began a recall of Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) and product codes 1M8560 and 1M8560R (pumps without drug... on . The reason FDA records is "The occlusion detection and alarm may be delayed when the pumps are used with larger volume syringes at low flow rates". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0599-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) and product codes 1M8560 and 1M8560R (pumps without drug library option); Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The occlusion detection and alarm may be delayed when the pumps are used with larger volume syringes at low flow rates.
- FDA root cause
- Other
- Action
- Important Product Information letters dated 2/8/06 were sent to the AS40 customers via first class mail to the attention of the Director of Nursing. The accounts were informed of the delay in the detection of a downstream occlusion when the AS40 pump is used with large volume syringes at low flow rates. To minimize the occlusion detection times at low flow rates, Baxter recommended that the customers take the following actions: * Use the smallest possible syringe size for the volume of fluid to be delivered; * Select the low occlusion pressure setting to minimize the detection time; * Utilize infusion sets with thick-walled, small bore tubing to minimize the detection time; and * Always verify that the clamping device or stopcock is open in the fluid path prior to starting an infusion.
- Quantity in commerce
- 17,400 pumps
- Distribution
- Nationwide and Canada
- Product codes and lots
- all serial numbers