Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication...
Baxter Healthcare Corp. began a recall of Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A on . The reason FDA records is "A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0612-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.
- Product code
- FRN Pump, Infusion
- Reason for recall
- A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Baxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply form confirming receipt of the letter. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
- Quantity in commerce
- 282 pumps
- Distribution
- Nationwide. Arizona, California, Florida, Georgia, Hawaii, Iowa, Indiana, Kentucky, Louisiana, Michigan, Mississippi, North Carolina, Nebraska, New Mexico, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, Washington, West Virginia
- Product codes and lots
- serial numbers 15060001AB, 15060002AB, 15060004AB/15060013AB, 15060015AB/15060021AB, 15060023AB/15060073AB, 15060075AB, 15060077AB/15060116AB, 15060116AB, 15060118AB/15060126AB, 15060128AB/15060130AB, 15060132AB/15060172AB, 15060174AB/15060209AB, 15070003AB/15070006AB, 15070008AB/15070018AB, 15070021AB, 15070023AB, 15070026AB/15070034AB, 15070037AB15070041AB, 15070044AB/15070047AB, 15070049AB/15070052AB, 15070056AB/15070059AB, 15070066AB, 15070069AB, 15070071AB, 15070072AB, 15070076AB, 15070077AB, 15070079AB, 15070081AB, 15070082AB, 15070084AB, 15070086AB, 15070087AB, 15070092AB, 15070094AB/15070096AB, 15070099AB, 15070101AB, 15070109AB, 15070111AB/15070115AB, 15070117AB, 15070119AB, 15070121AB, 15070122AB, 15070124AB, 15070126AB/15070128AB, 15070130AB/15070132AB, 15070138AB/15070140AB, 15070144AB