Z-0604-06
FreeStyle Connect Point of Care Blood Glucose Monitoring System, including Control Solutions and Test Strips...
Abbott Medical / initiated 2006-02-17 / Terminated
Abbott Diabetes Care, Inc. began a recall of FreeStyle Connect Point of Care Blood Glucose Monitoring System, including Control Solutions and Test Strips; Part Number: 70672-01; Recalling... on . The reason FDA records is "The device may give inaccurate elevated blood glucose test results. This may lead to over treatment and hypoglycemic complications may result". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0604-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- FreeStyle Connect Point of Care Blood Glucose Monitoring System, including Control Solutions and Test Strips; Part Number: 70672-01; Recalling Firm/Manufacturer: Abbott Diabetes Care, Inc., 1360 South Loop Road, Alameda, CA 94502
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- The device may give inaccurate elevated blood glucose test results. This may lead to over treatment and hypoglycemic complications may result.
- FDA root cause
- Other
- Action
- On 2/17/06 the firm initiated the recall via its reps and on 2/22/06 notification was via letters explaining the reason for the recall and reporting that its reps will replace the units.
- Quantity in commerce
- 900
- Distribution
- The product was released for distribution to 28 consignees in US. The recall was appropriately extended to the consumer/user level; i.e., the physicians and hospitals who received the recalled product. There were no shipments to U.S. Government facilities or foreign or Canadian distribution.
- Product codes and lots
- All lots, including: FC0010, FC0012, FC0013, FC0014, FC0015, FC0016, FC0017, FC0020, FC0019, FC0018, FC0021, FC0022, FC0023, FC0026, FC0024, FC0025, FC0027, FC0028, FC0029, FC0030, FC0031, FC0032, FC0033, FC0034, FC0036