Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0855-06

CR DX-S, DX-S, Computed radiography system (Digitizer)

AGFA Healthcare Corp. / initiated 2005-11-03 / Terminated

AGFA Corp. began a recall of CR DX-S, DX-S, Computed radiography system (Digitizer) on . The reason FDA records is "Corrupted image after system start; Corrupted image after workflow interruption; An erasure unit problem can cause system to stop". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0855-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
CR DX-S, DX-S, Computed radiography system (Digitizer)
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Corrupted image after system start; Corrupted image after workflow interruption; An erasure unit problem can cause system to stop.
FDA root cause
Other
Action
Consignees were notified at the time of upgrading. The software was upgraded on 11/3/2005, 12/19/2005 and 12/28/2005. A Mandatory Service Bulletin, No. 11, explaining the issues was shared with the accounts.
Quantity in commerce
3 units
Distribution
DC, MS, OR--three states
Product codes and lots
CR DX-S: Units with Serial numbers SN1001-SN1079, Software versions: STR1102B and below.