Z-0862-06

REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Control...

Konica Minolta, Inc. / initiated 2006-02-06 / Terminated / root cause: labeling and UDI

Konica Minolta Medical Imaging USA, Inc. began a recall of REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Control. Patient throughput optimization via offloaded... on . The reason FDA records is "If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0862-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Konica Minolta, Inc.
Product
REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Control. Patient throughput optimization via offloaded processing from various CR readers. Technologist review and audit. Key processing for specialized exams, totally automated processing for hands off operation. Centralized workflow device for dispersed readers environments. HIS/RIS patient status updates. Demographic Correction.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will populate the fields where no data has been entered with data from the previous patient.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
The firm phoned the customers on 2/6/2006 to inform them of the recall and the corrective action associated to the recall. This will be followed up with a fax, e-mail and letter. The final step will be visiting the location to install the correct software.
Quantity in commerce
58 units
Distribution
The units were distributed to hospitals nationwide through the firm''s sales force and resellers. There are no government accounts.
Product codes and lots
58 units distributed between June 2003 and January 2006. Software version prior to January 31, 2006.