SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound...
Sonosite, Inc. began a recall of SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound systems (MicroMaxx, TITAN, and SonoSite180PLUS)... on . The reason FDA records is "When SonoCalc IMT 3.0 software is used with the SonoSite 180Plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0713-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM SonoSite,Inc.
- Product
- SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound systems (MicroMaxx, TITAN, and SonoSite180PLUS). This software generates a report with the patients intima media thickness (IMT) based on average thickness of the carotid artery.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When SonoCalc IMT 3.0 software is used with the SonoSite 180Plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events. Underestimates percentage of stenosis, false negative result.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- On October 26, 2005, the company sent their customers a User Guide Addendum that told customers how to verify that the image calibration data was correct.
- Quantity in commerce
- 155 units of the SonoCalc product to about 80 customers/distributors
- Distribution
- Nationwide, Kuwait, Mexico, South Africa, Chile, Spain, Japan, Puerto Rico, Australia. A list of 155 cases in which the software was sent out lists medical facilities, physicans, and SonoSite Distributors.
- Product codes and lots
- All SonoCalc IMT version 3.0