Z-0384-2012
Product is 65mm Flexible Tube***NON-STERILE PRODUCT***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY...
Biomet Microfixation / initiated 2006-02-17 / Terminated / root cause: design
Walter Lorenz Surgical, Inc. began a recall of Product is 65mm Flexible Tube***NON-STERILE PRODUCT***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN." W.LORENZ... on . The reason FDA records is "Lorenz Engineer discovered flex tube could break if bent too much". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0384-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biomet Microfixation
- Product
- Product is 65mm Flexible Tube***NON-STERILE PRODUCT***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN." W.LORENZ SURGICAL***WALTER LORENZ SURGICAL, INC. 1520 TRADEPORT DRIVE , REPLY TO POST OFFICE BOX 18009, JACKSONVILLE, FLORIDA 32229-8009 USA, 1-800-874-7711. Part # 80-3112-01 Lot 976580 Qty 41 10/18/2005. Sold for use with: Part # 80-3112-01 Lot # 976580 Qty 28. Part # SP-1652 Lot # 048720 Qty 3. Part #-1640 Lot # 056620 Qty 8. Part # SP-1651 Lot # 056590 Qty 2. Part # 80-3112-01 Lot 970060 Qty 31 10/20/2005. Sold for use with: Part # SP-1651 Lot # 056590 Qty 8. Part # SP-1653 Lot # 056610 Qty 10. Part # SP-1285 Lot # 056570 Qty 4. Part # SP-1652 Lot # 056600 Qty 9. Product Usage: Distraction devices are for use in bone stabilization and elongation (lengthening) when correction of oral, cranial, mandibular and maxiollofacial deficiencies or post traumatic defects require gradual bone distraction. Each implantable distraction device has a titanium alloy drive screw mechanism and LactoSorb resorbable correction plates, which are implanted by fixation with LactoSorb bone screws.
- Product code
- JEY Plate, Bone
- Reason for recall
- Lorenz Engineer discovered flex tube could break if bent too much.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- W. Lorenz Surgical notified consignees of recall by an Urgent Advisory Notice letter. Consignees were requested to acknowledge receipt of recall and provide product disposition. Initial letter was sent on March 2, 2006, March 15, 2006, follow-up letters sent on April 4, 2006 and May 15, 2006. For additional information call (904) 741-4400 or fax (904) 741-4500.
- Quantity in commerce
- 41 pieces and 31 pieces respectively
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: GA and TX and the country of Japan.
- Product codes and lots
- Lot # 976580 and Lot # 970060
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.