Z-0767-06

STS-400 Level1 Skin Temperature Sensor with 400 Series Thermistor

Abbott Medical / initiated 2006-03-08 / Terminated

Smiths Medical ASD, Inc. began a recall of STS-400 Level1 Skin Temperature Sensor with 400 Series Thermistor on . The reason FDA records is "Product temperature readings may be 2C higher than the actual temperature". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0767-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
STS-400 Level1 Skin Temperature Sensor with 400 Series Thermistor
Product code
FLL Continuous Measurement Thermometer
Reason for recall
Product temperature readings may be 2C higher than the actual temperature
FDA root cause
Other
Action
Smiths Medical notified consignees by letter dated 3/8/05. Customers are instructed to complete customer reply notices and return affected products for replacement.
Quantity in commerce
20,000
Distribution
Nationwide, Canada, France, Japan, Netherlands
Product codes and lots
Lot Number: 3141