Z-0856-06
Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667
B & K Medical Systems, Inc. / initiated 2006-03-16 / Terminated / root cause: manufacturing process
B & K Medical Systems, Inc. began a recall of Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667 on . The reason FDA records is "Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0856-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B & K Medical Systems, Inc.
- Product
- Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- BK Medical notified customers by letter on 4/10/06 via FedEx advising user of the problem and scheduling the software correction by a Service representative visit .
- Quantity in commerce
- 28 units
- Distribution
- CA,CT, FL, MD, ME, MN, MS, NY, OH, RI, SD, TX, VA
- Product codes and lots
- Serial Numbers: 1855380, 1855547, 1855548, 1856835, 1856837, 1857171, 1857173, 1857180, 1857427, 1857435, 1857515, 1857748, 1858482, 1858841, 1859751, 1859902, 1860250, 1860262, 1860487, 1860661, 1860858, 1861488, 1861580, 1861581, 1861823, 1861849, 1863084, 1863082