Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667

B & K Medical Systems, Inc. / initiated 2006-03-16 / Terminated / root cause: manufacturing process

B & K Medical Systems, Inc. began a recall of Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667 on . The reason FDA records is "Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0856-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B & K Medical Systems, Inc.
Product
Convex Array Transducer used with B-K Ultrasound Scanners Model number: 8667
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Due to a problem with calibration, the device can produce up to 50% elevated acoustic output and up to 2 degrees higher surface temperature
Root cause tag
manufacturing process
FDA root cause
Other
Action
BK Medical notified customers by letter on 4/10/06 via FedEx advising user of the problem and scheduling the software correction by a Service representative visit .
Quantity in commerce
28 units
Distribution
CA,CT, FL, MD, ME, MN, MS, NY, OH, RI, SD, TX, VA
Product codes and lots
Serial Numbers:  1855380, 1855547, 1855548, 1856835, 1856837, 1857171, 1857173, 1857180, 1857427, 1857435, 1857515, 1857748, 1858482, 1858841, 1859751, 1859902, 1860250, 1860262, 1860487, 1860661, 1860858, 1861488, 1861580, 1861581, 1861823, 1861849, 1863084, 1863082