ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit...
Abbott Laboratories MPG began a recall of ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay... on . The reason FDA records is "The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0860-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay diluent, pre-treatment 1, pre-treatment 2, and pre-treatment 3; Abbott Laboratories, Abbott Park, IL 60064
- Product code
- LIG Radioassay, Intrinsic Factor Blocking Antibody
- Reason for recall
- The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT instrument.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Abbott affiliates in Canada, Germany and Australia were e-mailed copies of the recall letter on 3/8/06 for translation where needed and dissemination to their accounts receiving the suspect lots. The end users were advised that the 30-day onboard storage information is not included in the barcode for the two lots. As a result the ARCHITECT system software is unable to track how long the reagent kit has been stored onboard. The accounts were instructed to manually track the onboard storage time for the two lots, and discard the kits after 30 days of onboard storage.
- Quantity in commerce
- 135 kits
- Distribution
- Canada, Germany and Australia
- Product codes and lots
- lot numbers 33079M200, 33079M201