Z-0860-06

ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit...

Abbott Medical / initiated 2006-03-08 / Terminated / root cause: labeling and UDI

Abbott Laboratories MPG began a recall of ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay... on . The reason FDA records is "The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0860-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay diluent, pre-treatment 1, pre-treatment 2, and pre-treatment 3; Abbott Laboratories, Abbott Park, IL 60064
Product code
LIG Radioassay, Intrinsic Factor Blocking Antibody
Reason for recall
The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT instrument.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Abbott affiliates in Canada, Germany and Australia were e-mailed copies of the recall letter on 3/8/06 for translation where needed and dissemination to their accounts receiving the suspect lots. The end users were advised that the 30-day onboard storage information is not included in the barcode for the two lots. As a result the ARCHITECT system software is unable to track how long the reagent kit has been stored onboard. The accounts were instructed to manually track the onboard storage time for the two lots, and discard the kits after 30 days of onboard storage.
Quantity in commerce
135 kits
Distribution
Canada, Germany and Australia
Product codes and lots
lot numbers 33079M200, 33079M201