Baxter Healthcare Corporation, Renal Software Suite v.3.1 - HD Module; clinical data management software...

Baxter Healthcare Renal Div / initiated 2006-03-29 / Terminated / root cause: software

Baxter Healthcare Renal Div began a recall of Baxter Healthcare Corporation, Renal Software Suite v.3.1 - HD Module; clinical data management software; Baxter Healthcare Corporation, McGaw Park, Illinois... on . The reason FDA records is "Software Error: There are software discrepancies that may increase the possibility of a medication error if the dialysis clinic software is used to track, record, or coordinate the administration of medications". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0833-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter Healthcare Corporation, Renal Software Suite v.3.1 - HD Module; clinical data management software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085
Product code
KPF System, Dialysate Delivery, Semi-Automatic, Peritoneal
Reason for recall
Software Error: There are software discrepancies that may increase the possibility of a medication error if the dialysis clinic software is used to track, record, or coordinate the administration of medications.
Root cause tag
software
FDA root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 3/29/06 to those customers who had their RenalSoft and Renal Software Suite software configured with the affected modules. The letters indicated in a table which software anomalies were associated with the affected software modules in each software version. A description of each anomaly was provided, along with interim actions to take to mitigate the anomaly until the new software version RenalSoft V.2.0.x can be issued. Individual customer training will be performed at each facility for those issues identified in the letter specific to their software version. At that time the mitigation provided in the letter will be reviewed with the customer.
Quantity in commerce
6 - U.S., 44 - International
Distribution
Nationwide (Alabama, Arizona, California, Colorado, Connecticut, Illinois, Iowa, Kansas, Maryland, Massachusetts, Michigan, Minnesota, Missouri, New York, Pennsylvania, South Dakota,Texas and Wisconsin); and internationally to Australia, Austria, Canada, Denmark, Finland, Greece, Ireland, Italy, the Netherlands, Norway, Portugal, Spain, Sweden and the United Kingdom.
Product codes and lots
Version 3.1 - HD Module only