Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0776-06

Roche MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog number 03264793001

Roche Diagnostics GmbH / initiated 2006-04-04 / Terminated

Roche Diagnostics Corp. began a recall of Roche MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog number 03264793001 on . The reason FDA records is "The firm has received complaints of reagent splattering on the deck of the instrument and possible cross contamination of samples purified on the instrument". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0776-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog number 03264793001.
Product code
NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr
Reason for recall
The firm has received complaints of reagent splattering on the deck of the instrument and possible cross contamination of samples purified on the instrument.
FDA root cause
Other
Action
Consignees were notified by letter dated 4/4/06 to notify customers of the problem and provide a work-around.
Quantity in commerce
unknown.
Distribution
Nationwide.
Product codes and lots
All lots.